Appendix J
AccuPyc II 1340
J-4
June 2013
1- and 0.1-cm
3
Inserts
Special tools are required for installing the 1- and 0.1-cm
3
inserts; these tools are included in
the appropriate MultiVolume kits.
1-cm
3
Insert and Sample Cup
Perform the following steps to install the 1-cm
3
insert and sample cup:
1. Remove the sample chamber cap and place on a clean work surface with the greased side
facing upward. The cell chamber cap should be left off the sample chamber a minimal
amount of time to prevent contamination from airborne particles.
2. Place the insert into the sample chamber.
3. Use the handling tool to grasp the cup for placement in the insert. After placing the sample
into the sample cup, insert the tip of the tool into the cup and press down firmly. The
handling tool features an O-ring to grip the inside surface of the sample cup.
O-ring
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...