AccuPyc II 1340
System Commands
Aug 2013
3-13
Enables you to choose the manner in which to end the
pressure measurement.
Choices:
Equilibrate, Fixed interval
Press
CHOICE
to make your selection, then press
ENTER
.
Does not display for Fixed
Interval
Displays when you select
Equilibrate
at the
P1, P2 end by?
prompt; pressure measurement will end when the entered rate
is obtained. Enter the equilibration rate and press
ENTER
.
Range:
0.0001 to 9.0000 psig/min
0.0007 to 62.05 kPag/min
A high rate will produce faster results, but results may not be
as precise as desired. The lowest rates may cause errors when
some materials (such as those with appreciable vapor
pressures or organics) are analyzed.
Does not display for Equilibrate
Displays when you select
Fixed interval
at the
P1, P2 end
by?
prompt; pressure measurement ends when the specified
time is reached. Enter the interval time and press
ENTER
.
Range:
10 to 99999 s (seconds)
The run precision feature enables early termination of the
analysis when certain criteria are met.
Choices:
Yes, No
When you select
Yes
, the analysis terminates after five
consecutive cycles are within the specified tolerance. Always
request a large number (50 to 99) of runs. If you select a
small number of runs, the analysis stops when the number
you entered is reached even though the specified tolerance
has not been met.
Select
Yes
or
No
, then press
ENTER
.
Unit[n]
SN1234 10 cm3
Analysis Parameters
P1, P2 end by?
Equilibrate
Unit[n]
SN1234 10 cm3
Analysis Parameters
Equilibration Rate:
0.005 psig/min
Unit[n]
SN1234 10 cm3
Analysis Parameters
Interval Time:
60 s
Unit[n]
SN1234 10 cm3
Analysis Parameters
Use Run Precision?
No
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...