Appendix A
AccuPyc II 1340
A-2
Oct 2013
ANLSERR: Eq failure
NN/ZZ cycles completed
ANLSERR: Fill failure
NN/ZZ cycles completed
ANLSERR: Overrange
NN/ZZ cycles completed
Cause A:
The sample being analyzed failed to equilibrate in 1000 seconds.
Action A:
Check the system carefully for leaks as described in Checking the Cell
and Expansion Chambers for Leaks later in this chapter. Make sure the
sample is properly prepared before performing an analysis.
Cause B:
Sample (foams, organics, etc.) absorbs helium slowly.
Action B:
Set the equilibration rate to progressively higher values until
reasonable equilibration times (15 sec to 120 sec) are achieved.
Cause:
There was insufficient pressure to allow filling within five minutes
during an analysis.
Action:
Increase the regulator pressure or, if there is insufficient helium in the
tank, obtain a new tank of helium.
Cause A:
A pressure overrange occurred during an analysis because the regulator
pressure is set too high.
Action A:
Adjust the pressure so that it is equal to the higher of the two fill
pressures specified in the analysis parameters plus 2.0 psig.
Cause B:
A pressure overrange occurred during an analysis because an error
occurred in the pressure measurement electronics.
Action B:
Call a Micromeritics Service Representative.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...