Scala2 | Version 5 | Maintenance and Troubleshooting
• Active wideband hydrophone with high/low gain: below -100 dBV
• Active narrowband: NC-1-04 below -80 dBV / NC-1-07 below -100 dBv
• Passive hydrophone: below -110 dBV
6.
Scroll on the frequency or dBV scales to zoom in and out.
7.
Under
Peak
, you can check:
•
RealTime
: the latest highest level of noise (dBV)
recorded and its frequency.
•
Max
: the highest level of noise recorded since the
beginning of the spectrum and its frequency.
8.
Check that there is more than 12 dBV between the maximum noise level (dark blue line) and the
average noise level (cyan line) on the peak of sensor frequencies.
9.
If you changed the configuration of the hydrophone or sensors, right-click the graph and click
Reset Max
to reset the dark blue line showing the maximum level of noise.
10.
To check the maximum, mean and real time measures of noise level at specific frequencies:
a)
Right-click the FFT plot and click
FFT Marker
.
b)
Click and drag the marker at a specific point.
Frequency and levels of noise at the marker position are displayed on the right side of the
graph.
11.
Right-click the spectrum and click
Pause
if needed.
12.
To save data recorded by the spectrum in a *.txt file, right-click the FFT plot and click
Save
FFT
.
|
152
Summary of Contents for Scala2
Page 1: ...Scala2 User Manual...
Page 2: ......
Page 20: ...Scala2 Version 5 Installation 19...
Page 28: ...Scala2 Version 5 System Configuration Single Trawl Gear Twin Trawl Gear Triple Trawl Gear 27...
Page 29: ...Scala2 Version 5 System Configuration Quadruple Trawl Gear Twin Outrigger Trawl Gear 28...
Page 30: ...Scala2 Version 5 System Configuration Port Seine Gear Starboard Seine Gear 29...
Page 69: ...Scala2 Version 5 System Configuration 68...
Page 105: ...Scala2 Version 5 Display Configuration 104...
Page 122: ...Scala2 Version 5 Display Configuration 121...
Page 156: ...Scala2 Version 5 Maintenance and Troubleshooting c Double click UninstallScala command 155...