7.10.4 Defrosting/cooldown after transport/storing in temperatures below 5°C or above
40°C
If the Innowalk has been transported or stored in temperatures below +5°C, the product must be
defrosted before use. Low temperatures may cause failure in switches for positioning and cause the
mechanisms to move heavier than normal.
If the Innowalk has been stored or transported in temperatures colder or warmer than the running
temperature, it should be given 24 hours to reach ambient temperature.
Temperature Running
o
C
+5 - +40
Storing/transport
o
C
-40 - +70
7.10.5 Hot surfaces
Innowalk has no hot surfaces under normal use except the motor casing that can get warm during
use. Do not hold on the motor casing during use.
During charging the USB-charger on the front tilt can get warm to the touch and is to be expected.
During use in high ambient temperatures close to 40
o
C the motor cover, remote control, USB-
Charger and control panel buttons can achieve temperatures up to 52
o
C. When the product is used
under these conditions, remove your hand from the buttons on the remote control, plug and control
panel when the operations is performed to avoid warm hands and/or discomfort. Do not be in
contact with the remote control and control panel for more than 10 minutes continuously under
these conditions. Do not be in contact with the USB-charger more than 1 minute
if it’s hot to the
touch.
Using the product in high ambient temperatures should be assessed regarding the user, to prevent
unwanted situations that can arise from activity in high temperature (eg: Fatigue, heat stroke,
dehydration, concentration issues etc (this list is not exhaustive).
Important:
If any other surfaces at the Innowalk become hot during use, or the cover,
remote control, USB-charger or control panel buttons become so warm it causes discomfort
to touch them, then immediately stop the use and pull out the Innowalk power cord from the
wall outlet. This connection is found on the control panel on the side of the motor cover,
where the power cord is attached to the Innowalk
. The “ON” indicator light should
then dim
and stop when the power cord is removed from the power outlet. Contact your Made for
Movement representative to examine the product before using again.
Picture 58
–
AC plug