Users manual for Innowalk Pro 2
Version: 1.06
–
Version date: 06.12.2023
Page
62
of
78
Fuse:
Note: No adjustment, not described in this users manual, should be performed by any other than a
Made for Movement representative who have received training in adjusting the Innowalk.
Speed and Power:
Speed range: 10-85 RPM (0-60 RPM is the most used range: default)
Acceleration, 0 to 60 RPM: 10 sec. (Stepless)
Deceleration, 60 to 0 RPM: 0,3 sec. (Instantly. Pressing the Stop button at full speed)
Deceleration, 60 to 10 RPM: 10 sec. (Holding the slower button from max speed, until slowest speed
is reached. Stepless)
Power: 200 Watt motor output
Acceleration/deceleration/speed rates can be changed by MFM if necessary.
Spasm Control:
Please see chapter 5.3 for details regarding adjustment of the spasm control, and who can adjust the
spasm control.
Sounds:
The verification tone from remote control button pushes is approx. 50 db.
The sound from the unit during use is normally between 50 and 52 db.