Users manual for Innowalk Pro 2
Version: 1.06
–
Version date: 06.12.2023
Page
61
of
78
Shoe clip
kg
9,1
Ski
kg
3,4
Seat and upright
kg
37,5
Back support
kg
12
Rear module
kg
60
Arm-moving tubes w.brackets kg
6,5 (2pcs)
Main power supply Innowalk
Power-supply, input
AC 110-230V 50
–
60 Hz
6.0
–
2.6A
Power-supply, output DC 24V
20.8 amp.
500W
Most relevant momentary ratings: Volt = 24 V ampere, 10 amp.
Motor
DC 24V
200W
USB Charger on front tilt: Input 24V. Output: 5,2V, max 2,5 amp.
The USB charger can be used to charge phones/tablets. The Innowalk is a stand-alone medical
equipment (ME), where the attachment of phone/tablet is not considered nor evaluated as part of
the ME system. The USB is for charging and has no other function than that.
Temperature* running
o
C
+5 - +40
Storing/transport
o
C
-40 - +70
Relative humidity*
running
%
35-70
Storing/transport
%
15-93
Atmospheric pressure
kPA
70
–
106
* See chapter 7.10.4 for defrosting and warmup before use
Operating time: Max 1 hour continuous use on each session.
Cooldown/pause: Minimum 10 minutes equipment pause needed between each use session.
Use category/classification: Non-continuous operation.
IP degree: IP22
Expected service life at main
product, accessories, remote
control and power supply:
5 years