Users manual for Innowalk Pro 2
Version: 1.06
–
Version date: 06.12.2023
Page
52
of
78
Accessible parts:
The below parts will or can come in contact with the operator during normal use.
Accessible parts are defined to usually be in contact with the user and/or operator less than a minute
at the time under normal conditions. Contact with the parts usually occur in connection with
transfer, attachment and use of the product with the user. Accessible parts are:
•
All parts necessary for adjustment of the Innowalk
•
Padding, straps, velcro
•
Most metal parts not covered by the motor cover
•
The remote control and control panel buttons
•
Stickers and labels
•
The power unit, with power cable. The power unit can get warm during use. Do not hold on
the power unit/cover during use.
7.7
Shoes
Most shoes can be worn with Innowalk. Make sure the shoes are tightened well to the users foot, so
the user is not able to pull his/her foot out of the shoe while legs are moving in the Innowalk.
Ankle-foot orthosis may be used in the Innowalk, if the type of orthosis and limitations related to
range of motion of the ankle joint is taken into consideration during individually assessment of the
user by the coordinator.