4
Set up the Counter
Overview
This chapter duplicates and consolidates information provided
elsewhere in this manual as a convenience to the reader.
Analog
Sensors
The SOLAIR can be fitted with up to four 4-20mA Analog sensors.
The Sensors are connected to the instrument using RJ-12 cables. The
connector pinout is shown in Table 4-1.
Note:
Changes to the
analog configuration will
change historical
environmental values in
the data buffer.
Each port can be used for one dual-channel or two single-channel 4-
20mA sensors (when wired correctly), two devices total per port. When
each port is connected to two single-channel analog sensors, a total of
four analog sensors may be connected to the SOLAIR. Otherwise, only
one single-channel or one dual-channel device may be used at a time on
each port. Please contact Lighthouse for additional information about
converters and wiring.
Table 4-1 Analog Connector Pinout
Pin
Number
Function
1
5VDC
2
Analog #1 In
3
24VDC (VCC)
4
GND (tied to Pin 6)
5
Analog #2 In
6
GND
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......