SOLAIR 3350, 5350 MODBUS Register Map v1.48
248083387-1 Rev 10
B-11
Data Type Registers
Note:
All data records
have the same data types
assigned to them. The
user does not have to
read the data type
registers for every
record.
The 41xxx register series is used to identify the type of data items in the
30xxx series. The Data Type registers run in parallel with the Data
Registers. For example, Data Register 30041’s Data Type register is
41041.
Data Types are assigned 4 ASCII characters across 2 registers. If a
Data Type string contains less than 4 characters, then the rest of the
string is padded with NULL characters. Note that a Data Type using all
four characters will not end with a NULL character.
Table B-10 Data Types
String
Description
TIME
Timestamp
STIM
Sample Time
SVOL
Sample Volume
LOC
Location
STAT
Status
TEMP
Temperature
RH
Relative Humidity
AIRV
Air Velocity
DPRS
Differential Pressure
ESD
Electrostatic Discharge
FLOW
Flow Rate
LASV
Laser Voltage
VOLT
Voltage
PRES
Pressure
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......