Setting up Counter to Run Reports
For this example, the FED STD FT
3
report will be used. See
Figure 6-93.
Figure 6-93 FED STD FT3 Report Screen
•
Set the instrument for the Fed Std ft
3
report, 100 ft
2
, Class 100,
unidirectional air flow, operational room status.
•
This will generate the following minimum requirements:
Note:
The Minimum
Volumes are measured in the
standard measurement
required by each specific
report.
The number of digits allowed
for area to calculate volume is
limited to a 5 digit value for ft2
and a 4 digit value for m2.
•
Min Locations: 4
•
Min Samples: 5
Table 6-1 Table - Minimum Volume per Channel Size
Ch Size (
P
)
Min Vol (ft
3
)
0.3
0.100
0.5
0.200
1.0
0.917
3.0
10.309
5.0
31.696
10.0
145.985