Lighthouse SOLAIR 3350, 5350 Gen E Operating Manual
6-38
248083387-1 Rev 10
For RS-485 communications, each device on a multi-port chain must
have a unique address. See Figure 6-49.
Figure 6-49 COMM Address Configuration screen
Set the COMM Address by using the numeric keypad to select the
address; press ERASE to erase a number, if needed. Press ENTER to
accept the value. Press BACK to return to the Configuration screen or
press MAIN to return to the MAIN screen.
Output Setup
The Output Setup screen provides the option to send reports to a USB
flash drive or to the thermal printer. If USB is chosen, the options are
limited to Output on Sample or Output on Alarm. The report can be
saved as either a CSV (Comma Separated Value) formatted text file or
an encoded LSD (Lighthouse Secure Data) non-human readable file
that can be imported and decoded in LMS Express. Figure 6-50
illustrates the USB options.