Operate Counter
248083387-1 Rev 10
6-55
Note:
When the buffer
has wrapped, the record
that is first displayed in the
Data screen is the first
record in the data buffer.
This may not be Rec#1.
If the single down arrow
button is pressed once, the
instrument will display the
last record in the data
buffer.
•
Rec#
- Identifies which record is currently viewed.
•
*Recs
- Displays how many records are currently stored in the
SOLAIR’s buffer. When there is an asterisk (*) next to the word
“Recs”, this means that the 3000 record data buffer has wrapped
and older data is being overwritten. The basic concept is First In,
First Out (FIFO).
•
Size
- Lists the channel sizes configured on the instrument.
•
Diff#, Diff/ft3, Diff/m3
- Indicates that each channel size’s data is
displayed in differential mode. If a channel was disabled, then there
is a blank space in is column; # indicates raw counts; ft
3
and m
3
indicate normalized counts.
•
Cuml#, Cuml/ft3, Cuml/m3
- Indicates that each channel size’s
data is displayed in cumulative mode. If a channel was disabled,
there will be a blank space in this column. # indicates raw counts
while ft
3
and m
3
indicate normalized counts.
•
Loc
- Indicates the location at which the data record was recorded.
The Location name listed is the alphanumeric label that was saved.
•
Smpl
- Indicates the Sample Time (HH:MM:SS) at which the data
record was sampled.
•
Date
- Indicates the instrument date on which the data was
recorded. The date will display in the format selected in the
configuration (MM/DD/YYYY, DD/MM/YYYY or YYYY/MM/
DD).
•
Time
- Indicates the instrument time at which the data was recorded
(HH:MM:SS).
•
Instr
- Indicates the state of the instrument at the time the data was
recorded. Instrument states include GOOD or SRVC.
If SRVC appears, printouts will say “Service Required” and the
sensor may need cleaning. Please contact Lighthouse Technical
Support at 800-945-5905 (USA Toll Free) or 541-770-5905
(Outside of USA).
•
Flow
- Indicates the flow state of the instrument at the time the data
record was recorded. Flow is recorded as OK or ALRM. If the flow
was in alarm, it will print as Flow: Alert on the printouts.
•
Alarm
- Indicates (NONE or YES) if the data record exceeded any
of the alarm thresholds of any of the channels that were enabled for
alarms. If the data record exceeded alarm thresholds, printouts
show “Alarm: Yes”.
•
Laser
- Indicates the status of the laser at the time the data record
was recorded; possible laser states include OK or SRVC. If the laser
needs to be serviced, it will print as Laser: Service.
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......