Lighthouse SOLAIR 3350, 5350 Gen E Operating Manual
6-36
248083387-1 Rev 10
the MAIN screen.
Note:
If the first channel
is disabled, the display and
zoomed data view will be
blank.
One Channel mode only affects how data is displayed on the MAIN
screen. When the instrument is in One Channel mode, data will
continue to be recorded, printed and downloaded for all channels. See
Figure 6-47.
Figure 6-47 Main Screen - One Channel Option enabled
ONE SECOND REALTIME MODBUS OUTPUT
If One Second Real-time MODBUS Output is enabled, the instrument’s
settings reset to the following.
Note:
Alarming is not
available when One
Second Real-time
MODBUS Output is
enabled.
•
Mode: AUTO
•
Cycles: Zero
•
Sample Time: One second
Note:
If any setting (mode,
sample time, hold time, etc.)
is changed, the One Second
Real-time MODBUS Output
mode will be automatically
disabled.
•
Hold Time: Zero seconds
•
Cumulative/Differential: Cumulative (CUML)
•
Raw/Normalized: Raw
When counting, the MAIN Screen will update continuously and data
will not be recorded to the data buffer.
Pump Startup
This is ON by default and sets a pump ramp up time at the beginning of
the first sample, or at the beginning of all samples if the HOLD time is
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......