Operate Counter
248083387-1 Rev 10
6-53
Figure 6-71 Auto Increment Prompt
•
The Auto Increment mode will select the next corresponding
location and continue to sample using the current configuration. If
the current location is using a recipe, the user will be prompted to
unload the recipe before moving to the next location.
Data View
Buffer Screen
Data stored on the instrument is viewed in the Data View Buffer. New
records overwrite the first records stored after the buffer fills to its limit
(3000 records). An asterisk (*) next to the word “Records” on the
MAIN screen indicates the Buffer is full and data has wrapped. The
first record shown in the Data View Buffer after data has wrapped will
be the most current record and not the first record stored. In this case,
older data will be lost if it hasn’t been printed or transferred to a PC or
management system.
Particle concentration data displays and prints in ft
3
if the Particle
Volume is set to ft
3
. If the Particle Volume is set to m
3
or Liters, data
displays and prints in m
3
.
Press the DATA button on the MAIN screen to display the Data View
Buffer.
Depending on the sample setting configuration, the data displays in
either RAW data mode as shown in Figure 6-72 or NORMALIZED to
ft
3
or m
3
as shown in Figure 6-73.
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......