Set up the Counter
248083387-1 Rev 10
4-21
10. Type
telnet xx.xx.xx.xx 9999
and press
Enter
. This command
accesses the telnet port of the instrument. The expected error
message from Step 9. is shown in Figure 4-18.
Figure 4-18 Command - Send “Telnet 9999"
The user will have 4 seconds to respond in the next step. If the
response is too slow, repeat step 10.
11. Press
Enter
to start the SOLAIR Ethernet interface setup
program. Figure 4-19 displays some of the default settings - make
no changes to these settings except as instructed.
Figure 4-19 Starting the Set Up Program
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......