7. For flow values, read registers 30067 and 30068 as Float32:
•
Flow value will be shown in CFM, for example:
•
Flow rate = 3.5 CFM displays 1;
•
Flow rate < 3.5 CFM displays .9;
•
Flow rate > 3.5 CFM displays 1.1
8. For laser values, read registers 30069 and 30070 as Float32:
•
Laser value is displayed as a percentage of the value set during
calibration of the instrument. For example, if the value during
calibration is 1.0V and the current value is 1.20V, the display
will show 120.000. If the current value is 0.5V, the displayed
value will be 50.000.