Operate Counter
248083387-1 Rev 10
6-27
ALARM
The Particle Alarm screen allows the user to enable alarming on
specific channels as shown in Figure 6-31.
Figure 6-31 Particle Alarm Configuration Example
Note:
Alarming is only
applicable for AUTO and
MANUAL mode. It applies
only to Raw particle counts
even if the instrument is
displaying Normalized
data.
If the instrument is set to
Differential data, the alarm
threshold will apply to the
differential counts.
If the instrument is set to
display Cumulative data,
the alarm threshold will
apply to the cumulative
counts.
To enable the alarming for any channel, press the “X” next to that
channel. When a checkmark is displayed, that channel is enabled for
alarming. Press the checkmark to disable the alarming for that channel.
Alarm Threshold
Press the threshold button next to the enabled channel in order to set the
alarm threshold for that channel. The Alarm Threshold screen will open
as shown in Figure 6-32.
Figure 6-32 Configure Alarm Threshold
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......