Lighthouse SOLAIR 3350, 5350 Gen E Operating Manual
7-6
248083387-1 Rev 10
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 7-4 Instrument Parameters
Register
Data Type
Description
40026
unsigned integer
Location number; Specifies location of
Particle Counter.
40029
unsigned integer
Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait
before starting the first sample. Max value
is 86,399, which equals 23h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait
between sample periods. Max value is
86,399, which equals 23h 59m 59s
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in
conjunction with 40034. Number of
seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Summary of Contents for SOLAIR 3350
Page 1: ......
Page 2: ...00...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Page 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Page 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Page 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Page 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Page 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Page 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Page 183: ...00...
Page 184: ......