Lighthouse SOLAIR 3350, 5350 Gen E Operating Manual
4-14
248083387-1 Rev 10
6. Select
Connections
and verify that
Connect Protocol
and
Connection settings are set to the following:
•
Protocol: Modbus
•
Local Port: 502
•
Accept Incoming: Yes
.
•
Under
Modbus Mode
, set the
Message Timeout to
“
1000ms
” if
not already set.
•
Press the
OK
button.Connections Settings will display
Done!
as
shown in Figure 4-11.
Figure 4-11 Connections