SOLAIR 3350, 5350
MODBUS Register
Map v1.48
COMM
Settings
Lighthouse particle counters using MODBUS require the following
communications settings:
The MODBUS slave address is set on the particle counter.
Supported
MODBUS
Commands
See www.modbus.org for documentation on how to use these
commands.
Table B-1 MODBUS Communications Settings
Baud Rate
19200
Data Bits
8
Stop Bits
1
Parity
None
Hardware Protocol
RS485, USB and Ethernet
Software Protocol
MODBUS ASCII (supports upper/lower case)
MODBUS TCP
Table B-2 MODBUS Registers
Hex Command
Description
03
Read Holding Registers
04
Read Input Registers
06
Write Single Holding Register