![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 86](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367086.webp)
Getting Started Manual
WP7Zi-GSM-E RevB
86
W
IDTH
C
ONDITION IS
Width triggers have additional Width Condition settings as follows.
Your
Width Condition
can be specified as
Less Than
,
Greater Than
,
In
Range
and
Out Of Range
.
In Range
and
Out Of Range
conditions can be further set with
Upper
and
Lower Values
(for
Limits
).
OR
Nominal width
and
Delta
(for
Delta
).
Note
: Width Condition is settings are summarized to the right of the
dialog.
H
OLDOFF BY
T
IME OR
E
VENTS
Holdoff is an additional condition of Edge and Pattern triggers. It can be
expressed either as a period of
time
or an
event
count. Holdoff disables
the trigger circuit for a given period of time or number of events after the
last trigger occurred. Events are the number of occasions on which the
trigger condition is met. The trigger resumes when the holdoff has elapsed
and its other conditions are met.
Select the holdoff by touching the
Time
or
Events
buttons on the
Holdoff
dialog.