![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 38](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367038.webp)
Getting Started Manual
WP7Zi-GSM-E RevB
38
Quick Zoom
- Press to automatically display magnified views of up
to four signal inputs on multiple grids. With four input signals, the
signals are displayed along with four zoom traces, each on its own
grid. Pressing this button also turns off all other traces.
Vertical Position
- Press to reset the vertical zoom position to zero.
Turn to change the vertical position of the selected math or zoom
trace.
Vertical Ratio
- Press to toggle between fixed and variable vertical
zoom ratio adjustment. Turn to change the vertical zoom ratio of
the selected math trace.
Connecting
Probe Interfaces
LeCroy Zi Series oscilloscopes contain probe interfaces that provide a
complete measurement solution from probe tip to oscilloscope display. All
probe interfaces permit automatic recognition of connected probes. For
active single-ended and differential voltage probes and current probes,
these interfaces upload gain and offset correction factors from the probe
EEPROM's and automatically compensates to achieve fully calibrated
measurements.
This intelligent interconnection between your instrument and a wide range
of accessories offers important advantages over standard BNC and probe
ring connections. The system ensures correct input coupling by auto-
sensing the probe type, thereby eliminating the guesswork and errors that
occur when attenuation or amplification factors are set manually.
Furthermore, it provides automatic matching of probe to oscilloscope
response - ensuring the most pristine signal fidelity.
The LeCroy WavePro 7 Zi and Zi-A series oscilloscopes with bandwidths of
4 to 6 GHz include both ProBus and ProLink probe interfaces.