![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 60](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367060.webp)
Getting Started Manual
WP7Zi-GSM-E RevB
60
Refer to
Acquisition Modes
in the specifications on
www.lecroy.com
for
maximum sample rates.
C
OMBINING
C
HANNELS
P
ROCEDURE
Set up channel combinations on the interface as follows:
1.
Touch
Timebase → Horizontal Setup
on the menu bar and the
Horizontal Setup
dialog is shown.
2.
Under
Active Channels
, touch
4
,
2
or
Auto
. The maximum sample
rate is shown alongside each button.
Sampling Modes
Sampling Modes
Depending on your timebase, you can choose
Single-shot Sampling Mode
(below), also known as Real Time mode,
Sequence Sampling Mode
(on
page 61), or
RIS Sampling Mode
(on page 69) mode sampling.
Some instruments also offer
Roll Mode
(on page 70).
Selecting a Sampling Mode
1.
Touch
Timebase → Horizontal Setup
from the menu bar.
2.
In the
Timebase
dialog, touch a
Sample Mode
button.
3.
If you chose
Sequence Mode
, touch the
Sequence
tab. The
Sequence
dialog is shown. Use the following fields to provide details
for controls such as
Number of Segments
,
Enable Timeout
,
Timeout
(value),
Display Mode
, and
Show Sequence Trigger Times
.
Single-shot Sampling Mode
Basic Capture Technique
A single-shot acquisition is a series of digitized voltage values sampled on
the input signal at a uniform rate. It is also a series of measured data
values associated with a single trigger event.