WavePro 7 Zi and Zi-A Oscilloscopes
169
WP7Zi-GSM-E RevB
Touch Screen Calibration
Start the calibration procedure by touching the
Touch-Screen Calibration
button.
During calibration you are prompted to touch the center of a small cross in
five key touch screen locations.
PLEASE NOTE THE FOLLOWING
:
Use a stylus for this procedure. Using your finger for the calibration
prompts provides insufficient accuracy for the procedure.
The calibration procedure times out after ten seconds pass and no
cross is touched.
Avoid parallax errors by placing your line of sight directly in front of
each cross before touching it.
Managing Files
Use the instrument's
Utilities
menu to save the screen image as a file to
storage media, such as a USB drive or hard drive. You can give your files
custom names and create directories for them.
Hard Disk Partitions
The instrument's hard disk is partitioned into drive
C:
and drive
D:
. Drive C:
contains the Windows operating system and the instrument application
software. Drive D: is intended for data files.
System Recovery Tool
Restoring Software Using the Acronis Application
Your oscilloscope is designed to operate reliably for many years. The
application software operating the instrument runs on a Windows
platform. The loading or incomplete removal of additional Windows
applications can cause problems in the stability of the operating system.