![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 133](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367133.webp)
WavePro 7 Zi and Zi-A Oscilloscopes
133
WP7Zi-GSM-E RevB
If you choose either
WithinDeltaPct
or
WithinDeltaAbs
from the
Condition
menu (respectively, previous), you also have the choice of
setting the limit by means of the statistical buttons at the bottom of
the
ParamCompare
dialog:
Comparing Dual Parameters
1.
Touch
Analysis → Pass/Fail Setup...
from the menu bar.
2.
Touch one of the
Qx
tabs and a setup dialog for the position is
shown.
3.
Touch inside the
Condition
field in the main dialog and select
DualParamCompare
.
4.
Touch inside the
Source1
and
Source2
fields and select a source
from the pop-up menu.
5.
Touch inside the
ParamCompare
mini-dialog field and select a
source from the pop-up menu.
6.
Touch inside the
Compare Values
field and select
All
or
Any
from
the pop-up menu.
By selecting
All
, the test are true only if every waveform falls within
the set limit. Selecting
Any
makes the test true if just one waveform
falls within the limit.