WavePro 7 Zi and Zi-A Oscilloscopes
131
WP7Zi-GSM-E RevB
4.
Touch the
Summary View
checkbox to enable a line of text (as
follows) showing a concise status of the last waveform and keeping
a running count of how many sweeps have passed.
5.
Touch inside the
Pass If
field, and select a Boolean condition from
the choices shown.
6.
If you want to set up the test to end after a finite number of
sweeps, touch the
Stop Test
checkbox. Then touch inside the
After
data entry field and enter a value, using the pop-up numeric
keypad.
7.
Use either the
Pass
or
Fail
button on the
If
section of the dialog to
set actions taking place when your waveform passes or fails the
test.
8.
The
Then
section of the dialog provides
stop test
,
sound alarm
,
print result
,
emit pulse
, or
save
(the waveform) actions. If you want
to have the results printed (and your oscilloscope is not equipped
with a printer) make sure your instrument is connected to a local or
network printer. Learn more in the
Printing and Hardcopy
Functions
topic (available in the online help files).
9.
If you want to save your waveform automatically, touch the
Save
Setup
checkbox. This takes you out of the current dialog and opens
the
Save Waveform
dialog. Learn more in the
Save Recall → Saving
and Recalling Waveforms
topic (available in the online help files).
10.
Test Pass/Fail conditions by touching the
Force Actions Once
button. Press the
Clear All
button to quickly uncheck all checkboxes
and make new selections.
Comparing a Single Parameter
1.
Touch
Analysis → Pass/Fail Setup...
from the menu bar.
2.
Touch one of the
Qx
tabs and a setup dialog for the position is
shown.