Getting Started Manual
WP7Zi-GSM-E RevB
130
You can also decide the actions to occur upon your waveforms' passing or
failing, by selecting one or all of the following:
stop
audible alarm
print image of display
emit pulse (if available)
save waveform
The Pulse checkbox causes a pulse to be output through
the Aux Out connector at the front of the oscilloscope.
This pulse can be used to trigger another oscilloscope. You can set the
amplitude and width of the pulse as described in Auxiliary Output Signals.
Depending on your oscilloscope model, you can configure up to 8 pass/fail
conditions. The Boolean conditions to determine if your waveform passes
are as follows:
Pass/Fail Boolean Conditions
All True
All False
Any True
Any False
All Q1 to Q4
Or
All Q5 to Q8
Any Q1 to Q4
And
Any Q5 to Q8
P
ASS
/F
AIL
T
ESTING
S
ETUP
Initial Setup
1.
Touch
Analysis → Pass/Fail Setup...
on the menu bar.
2.
Touch the
Actions
tab.
3.
Touch the
Enable Actions
checkbox. This causes selected actions to
occur on your waveform's passing or failing a test.