Getting Started Manual
WP7Zi-GSM-E RevB
44
DC Level
- This is a reference level.
Trigger Enabled
- This function can be used as a gating function to
trigger another instrument when the oscilloscope is ready.
Pass/Fail
- Allows you to set a pulse duration from 1 µs to 131 ms;
generates a pulse when pass/fail testing is active and conditions are
met.
Aux Output Off
- This function turns off the auxiliary output signal.
Auxiliary Output Setup
1.
Touch
Utilities → Utilities Setup...
from the menu bar.
2.
Touch the
Aux Output
tab.
3.
If you want a 1 kHz, 1 V square wave, touch the corresponding
button.
4.
If you want a specialized output, touch one of the buttons under
Use Auxiliary Output For
.
5.
If you selected Square Wave, touch inside the
Frequency
data entry
field and enter a value, using the pop-up keypad. You can set a
value from 500 Hz to 5 MHz.
6.
If you selected Pass/Fail, touch inside the
Pulse Duration
field and
enter a value from 1 µs to 131 ms, using the pop-up numeric
keypad.