![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 135](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367135.webp)
WavePro 7 Zi and Zi-A Oscilloscopes
135
WP7Zi-GSM-E RevB
5.
From the
Test
portion of the dialog, make a selection in the
Test is
True when
group of buttons (as follows).
Note
: The previous selection is for
All In
and means the test is
False when a single waveform falls outside the mask.
6.
From the
Show Markers
section of the dialog, choose whether or
not to have mask violations displayed by touching either the
Off
or
On
button, respectively.
7.
If you are loading a pre-existing mask, touch the
Load Mask
tab, and
then the
File
button. You can then enter the file name or browse to
its location.
8.
If you want to make a mask from your waveform, touch the
Make
Mask
tab.
9.
Touch inside the
Ver Delta
and
Hor Delta
fields and enter boundary
values, using the pop-up numeric keypad.
10.
Touch the
Browse
button to create a file name and location for the
mask if you want to save it.
11.
Touch the
Gate
tab, then enter values in the
Start
and
Stop
fields to
constrain the mask to a portion of the waveform. Or, you can simply
touch and drag the Gate posts, which initially are placed at the
extreme left and right ends of the grid.
Customization
Customization Overview
The instrument provides powerful capability to add your own parameters,
functions, display algorithms, or other routines to the oscilloscope user
interface without having to leave the instrument application environment.