Power Consumption
</= 800 watts (800 VA) for all WavePro 7 Zi and Zi-A series models
(depending on accessories installed - probes, PC port plug-ins, etc.).
Power consumption in Standby Mode: 15 Watts.
Power and Ground Connections
The instrument is provided with a 10A/250V 18AWGrated grounded cord
set containing a molded three-terminal polarized plug and a standard
IEC320 (Type C13) connector for making line voltage and safety ground
connections. The AC inlet ground terminal is connected directly to the
frame of the instrument. For adequate protection against electrical shock
hazard, the power cord plug must be inserted into a mating AC outlet
containing a safety ground contact.
WARNING - Electrical Shock Hazard
Only use the power cord provided with your instrument.
Any interruption of the protective conductor inside or outside of the
oscilloscope, or disconnection of the safety ground terminal creates a
hazardous situation. Intentional interruption is prohibited.
In Standby mode, the oscilloscope is still connected to the AC supply. The
instrument can only be placed in a complete Power Off state by physically
disconnecting the power cord from the AC supply.
CAUTION
The outer shells of the front panel terminals (CH1, CH2, CH3, CH4, AUX IN,
AUX OUT) are connected to the instrument’s chassis and therefore to the
safety ground.