However, it is now possible to correct for group delay, and in doing so,
certain characteristics can be enhanced. By providing zero group delay at
all frequencies, preshoot and overshoot is equalized, which is preferable
for eye diagram measurements. Then, by providing the flattest possible
frequency response with zero group delay, the fastest possible rise time
can be measured. LeCroy provides all three of these choices in the
Response Optimization Mode selection.
Pulse Response
- A delay compensation that minimizes preshoot.
This selection most resembles the response of an analog
oscilloscope.
Eye Diagram
- Delay compensation with equalized preshoot and
overshoot. This selection improves the accuracy of serial eye
diagrams.