![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 176](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367176.webp)
Getting Started Manual
WP7Zi-GSM-E RevA
176
* In order to conform to Radiated Emissions standards, use properly
shielded cables on all I/O terminals.
** Meets Performance Criteria "B" limits during the disturbance; product
undergoes a temporary degradation or loss of function of performance
which is self recoverable.
† Meets Performance Criteria "C" limits during the disturbance; product
undergoes a temporary degradation or loss of function of performance
which requires operator intervention or system reset.
WARNING
This is a Class A product. In a domestic environment this product may
cause radio interference, in which case the user may be required to take
appropriate measures.
Low-Voltage Directive
EN 61010-1:2001
Safety requirements for electrical equipment for measurement, control,
and laboratory use.
The oscilloscope has been qualified to the following EN 61010-1 limits:
Installation Categories II (Mains Supply Connector) and I (Measuring
Terminals).
Pollution Degree 2 (Normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused by
condensation must be expected.)
Protection Class I (Provided with terminal for protective ground).
UL and cUL
The oscilloscope is UL and cUL listed and conforms to the following
standards:
UL Standard: UL 61010-1 2nd Edition.
Canadian Standard: CSA-C22.2 No. 61010-1-04.