![LeCroy WavePro 7 Zi series Getting Started Manual Download Page 31](http://html1.mh-extra.com/html/lecroy/wavepro-7-zi-series/wavepro-7-zi-series_getting-started-manual_1866367031.webp)
WavePro 7 Zi and Zi-A Oscilloscopes
31
WP7Zi-GSM-E RevB
Front Panel Groupings
The front panel is divided into sections based on functions. The following
sections explain them in a bit more detail.
Miscellaneous Setup Controls
Help
- Press to
open the
LeCroy Online
Assistant where
you can click to
open the oscilloscope online help table of contents, index, or search
for a topic using a keyword. If the second monitor is installed, the
online help opens on the second monitor.
Default Setup
- Press to reset the oscilloscope's settings to the
default configuration. Corresponds with screen menu selection:
File
→ Recall Setup → Recall Default Setup
. For a list of default settings,
see
Save/Recall → Saving and Recalling Setups
(on page 142).
Auto Setup
- Press once and the
Auto Setup
flyout menu opens.
OR
From the menu bar, touch the
Auto Setup
selection from the
Vertical
,
Timebase
, or
Trigger
menus.
Press the
Auto Setup
button on the flyout menu to perform a full
auto setup. Press a
Channel Find Scale
button on the flyout menu to
perform a quick auto setup for that channel only.