Appendix A
A-2
D960R3
User’s Manual Probot
To purge the dosage module:
a)
Switch the rotary valve into the PUMP position, remove the syringe (FIGURE
A-2) and push the piston all the way forward to remove any air or solvent
left. Refer to Section 3.3.3 for more information about how to replace the
syringe.
Waste
Fi
ll
Au
x
Reagent
Vials
Syringe Drive
Dosage Module
PUMP
FILL
PURGE
Rotary Valve
Needle
Tee
µ-Tee
OR
FIGURE A-2 Step 1 – Purge the Tubing
b)
Switch the rotary valve into the FILL position and withdraw the reagent
solvent from the vial. Fill the syringe completely and make sure that enough
reagent solvent is in the vial (FIGURE A-3).
Wa
st
e
Fill
Aux
Reagent
Vials
Syringe Drive
Dosage Module
PUMP
FILL
PURGE
Rotary Valve
Needle
Tee
µ-Tee
OR
FIGURE A-3 Step 2 – Fill the Syringe
c)
Switch the rotary valve into the PUMP position and purge the tube to the
needle (FIGURE A-4).
Summary of Contents for Probot 161403