Appendix C
C-2
D960R3
User’s Manual Probot
C.2
Table Alignment
To check the table alignment (X and Y), the Manual Control Unit (in conjunction
with the
µ
Carrier software) is used to manually move the table. Before
performing any check or aligning of the table, the Probot instrument must be
properly leveled.
C.2.1
Leveling
Before checking the alignment or performing any alignment, check to ensure that
the Probot is properly leveled. Use a spirit level if there is any doubt (FIGURE C-
10). If necessary, adjust the four feet until the instrument is leveled properly.
Front View
Right Side View
FIGURE C-10 Leveling the Probot
C.2.2
Test Needle and
µ
Carrier Setup
Note: The following procedure discusses all steps using the standard needle. If
the µ-Tee is installed, all steps are identical except step a).
A piece of fused silica is used as ‘test needle’. The table is then moved as close
as possible to the needle tip. The manual control unit is used to control the table
position.
a)
Install a piece of fused silica capillary. Adjust ‘X’ to 1 – 2 mm (0.04 – 0.08
inch).
Fused Silica Capillary
Tee
PEEK Adapter
Tube
Fitting
Needle
X
= 1 - 2 mm
FIGURE C-11 Installing the Fused Silica Capillary
Summary of Contents for Probot 161403
Page 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Page 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Page 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Page 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Page 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...