Appendix B
B-4
D960R3
User’s Manual Probot
To setup the fractionation task:
a)
Set the collection control mode to
External
on the
General
FIGURE B-4 General Settings of the Fractionation Task
b)
Enter the flow rate of the LC system and the calculated volume between the
detector flow cell and Probot needle outlet. The delay time will be calculated
automatically according to the flow rate of the LC system and the calculated
volume (FIGURE B-4).
c)
Define the
primary collect group
used for sample collection, select the
Collect
peaks only
option and define the waste group used to collect the mobile
phase in between the peaks on the
Destination
tab (FIGURE B-5).
FIGURE B-5 Destination Definitions of the Fractionation Task
Summary of Contents for Probot 161403