Appendix C
C-4
D960R3
User’s Manual Probot
c)
Move the table from one side to the other (X-direction) and check that the
needle is centered above the groove and that the distance to the needle tip
remains constant.
d)
Move the table to the front and to the back (Y-direction) and check that the
distance to the needle tip remains constant.
In case the table needs some adjustment, proceed as follows.
C.2.4
Alignment of the X-axis
To adjust the X direction:
a)
Loosen the two screws labeled ‘3’ in FIGURE C-13.
b)
Turn the two screws labeled ‘2’ in FIGURE C-13 clockwise to lift the table on
the left side, turn the screws counterclockwise to lower the table on the left
side.
Note: Turn both screws labeled ‘2’ in FIGURE C-13 by the same angle to keep
the Y-adjustment (e.g. to lift the table on the left side by 1mm, turn both screws
approximately 30 degrees clockwise).
c)
Retighten the screws labeled ‘3’ in FIGURE C-13 and check the X-direction
again.
d)
Repeat step a) – c) if necessary.
C.2.5
Alignment of the Y-axis
To check/adjust the alignment of the Y-axis:
a)
Remove the cover plate to get access to the adjustment screws (FIGURE C-
14).
FIGURE C-14 Removing the Cover Plate of the Y-axis
b)
Slightly unscrew the screws labeled ‘4’ – ‘6’ in FIGURE C-15, but do not
loosen screw 3.
Summary of Contents for Probot 161403
Page 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Page 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Page 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Page 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Page 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...