Appendix A
User’s Manual Probot
D960R3
A-3
Wa
st
e
Fill
Au
x
Reagent
Vials
Syringe Drive
Dosage Module
PUMP
FILL
PURGE
Rotary Valve
Needle
Tee
µ-Tee
OR
FIGURE A-4 Step 3 – Purge the Tubing
d)
Switch the rotary valve back into the FILL position, withdraw the reagent
solvent and fill the syringe completely.
e)
Remove the plunger from the syringe. Lower the syringe and allow the
contents to drain via gravity, then replace the plunger (FIGURE A-5).
Wa
st
e
Fill
Au
x
Reagent
Vials
Syringe Drive
Dosage Module
PUMP
FILL
PURGE
Plunger
(removed)
Rotary Valve
Needle
Tee
µ-Tee
FIGURE A-5 Step 4 – Drain the Syringe
f)
Reinstall the syringe.