D.1
Overview
This appendix describes the features and usage of the Vacuum Waste Removal
option. This option is provided as a factory installed option only. If your Probot
Micro Fraction Collector does not include this option, contact your local
representative or LC Packings for upgrade information.
The vacuum waste removal option allows the user to remove undesired portions
of the separation (‘waste’) without the need to spot them onto the target. A
vacuum that can be applied draws then the solvent into a waste container.
The vacuum waste removal option includes the following items:
•
Lift magnet
•
Vacuum pump
•
Waste container
•
Tubing and vacuum connectors
To control the vacuum waste removal, event outputs of the UltiMate capillary
HPLC system are used in combination with the peak detection capability of
CHROMELEON
®
(see also APPENDIX B).
The following sections provide information about how to install the vacuum
waste removal option and show how to setup the µCarrier and the
CHROMELEON software to control the waste removal option of the Probot. All
information provided refers to µCarrier version 2.0 and to CHROMELEON
6.5
SP3.
Note: The Vacuum Waste Removal Option can not be used in conjunction with
the dosage module (e.g. no matrix solution can be added).
Summary of Contents for Probot 161403
Page 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Page 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Page 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Page 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Page 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...