Display
Operating Procedure
Description
Programming Mode
“Technical” Programming
After accessing the “Technical” programming menu and entering the password, use the buttons
T1
and
T2
until the following screen is displayed.
Press the T3 button
to enter the menu.
Press the T3 button
to enter the menu, move between the parameters using the buttons
T1
and T2
to select
ENABLED
or
DISABLED
, press the T3 button
to confirm the option.
Press the T3 button
to enter the menu, move with the buttons
T1
and T2
to set the
desired temperature, press the T3 button
to confirm the value.
The first value indicates the time in seconds between the detection of the need to fill and the start of filling.
The second value indicates the time in seconds between filling and its actual end.
Press the T3 button
to enter the menu, move with the buttons
T1
and T2
to set the
desired automatic filling time, press the T3 button
to confirm the value.
• This parameter enables the technician
to set various parameters of the steam
boiler.
• The temperature of saturated water is
proportional to the pressure inside the
Steam Boiler. Therefore it is possible to
regulate the pressure of the steam boiler
by means of electronic temperature
control. Please use the following tables
as reference when setting the steam
boiler temperature.
Steam Boiler
Settings
Steam Boiler
Settings
Steam Boiler
Steam Boiler
ENABLED
Enter Password
******
Steam Temp.
123,7
°
C 123,5
°
C
Autofill Delay
2 start 2 stop
THE STEAM BOILER CONTAINS WATER AT ELEVATED TEMPERATURES.
WATER TEMPERATURE OVER 52°C CAN CAUSE SEVERE BURNS INSTANTLY OR DEATH FROM SCALDING.
DANGER
Temperature
Pressure
247°F/119°C
1.0 bar
260°F/127°C
1.5 bar
272°F/133°C
2.0 bar
The saturation temperature of water will vary
with ambient atmospheric pressure.
Tables in this manual are created for atmospheric
pressure at sea level. Slight adjustment in temperature
will be required for elevations other than sea level.
CAUTION
1
2
3
4
5