This chapter describes the commands listed in the overview of chapter C.2.
D.1 Instrument operational mode
The instrument should be put in maintenance mode while assembling or servicing the instrument. In maintenance mode the instru-
ment allows assembly and service activities and authenticates against fixed pre-set passwords.
The normal mode should be set when not assembling or servicing the instrument. The normal mode passwords will be kept in
maintenance mode and will be reactivated after switching back to normal mode.
D.1.1 Set instrument operational mode
Description
The command: ‘.set opermode=<mode>;’ will set the instrument operational mode.
Argument (1)
Description
The instrument operational mode.
Type
Text
string
Range
‘normal’,
‘maintenance’
Default
‘normal’
Example
Request:
.set opermode=maintenance;
Response:
.ok;
D.1.2 Show instrument operational mode
Descritpion
The command: ‘.show opermode;’ will show the instrument operational mode.
Argument (none)
Example
Request:
.show opermode;
Response:
.ok;
D.1.3 Clear user account (and operational mode)
Description
The command: ‘.clear account;’ will revert the user account passwords and operational mode to defaults. The command may be
used to go to a known state when logging in on one if the instrument interfaces fails.
D. Description of the commands
98
Summary of Contents for SOLYS Gear Drive
Page 1: ...SOLYS2 Sun Tracker SOLYS Gear Drive Sun Tracker Instruction Manual...
Page 2: ...2...
Page 4: ...4...
Page 6: ...6...
Page 10: ...10...
Page 39: ...Figure 3 13 Adjustment of the sun sensor 21 0 5 mm 3x 39...
Page 61: ...6 Insert the Ethernet cable Contacts on this side 61...
Page 78: ...78...
Page 80: ...80...
Page 86: ...86...
Page 88: ...88...
Page 112: ......