Drawings for both connectors and pin numbers including types of cable and number of strands are given in figure 2.9 and 2.10.
Figure 2.9. Power cable connector for 24 VDC
Figure 2.10. Power cable connector for 100 to 240 VAC
Please make sure the used cable is flexible enough to withstand the daily movement of the tracker.
2.6.2 Environmental conditions
Important,
SOLYS rear cover
SOLYS does not contain any user serviceable parts inside. Therefore it is advised not to open the rear cover. Doing so can defeat the
internal optical sensors which may result in physical damage to the SOLYS and/or any attached equipment. Such damage is not
covered under warranty.
Important,
Precipitation
With the SOLYS rear cover installed properly such that all 8 screws are secured, the SOLYS is environmentally protected to IP65,
according EN 60529: 1991 + CI 1993. This qualifies the SOLYS to be used in weather conditions with all types of precipitation.
Important,
Circuit breaker
A power on/off switch is not a part of the SOLYS; requirements for electrical equipment installed outdoors specify that:
• A power isolator (switch or circuit-breaker) must be included in the cable installation
• The isolator (switch or circuit-breaker) must be in close proximity to the equipment and within easy reach of the operator
• The isolator (switch or circuit-breaker) must be marked as the power disconnection device for the equipment used in weather
conditions with all types of precipitation.
21 mm
5 mm
9 to 12 mm
6 to 9 mm
Male connector (included)
Hirschmann: CA 3 LS
Cable connections:
1 = Red
(+24 VDC)
2 = Black (24 VDC GND)
= Green/Yellow (Protective Earth)
Power cable for outdoor use (not included)
Diameter: 6 to 12 mm
Conductors: 3
Flexible wires: 24 strands, diameter 0.2 mm for 0.75 mm2
32 strands, diameter 0.2 mm for 1 mm2
30 strands, diameter 0.25 mm for 1.5 mm2
Conductor area: 0.8 to 1.5 mm2
Conductor colors: Red, Black, Green/Yellow
21 mm
5 mm
Female connector (included)
Hirschmann: CA 3 LD
Cable connections:
1 = Brown (L)
2 = Blue
(N)
= Green/Yellow (Protective Earth)
Power cable for outdoor use (not included)
Diameter: 6 to 12 mm
Conductors: 3
Flexible wires: 24 strands, diameter 0.2 mm for 0.75 mm2
32 strands, diameter 0.2 mm for 1 mm2
30 strands, diameter 0.25 mm for 1.5 mm2
Conductor area: 0.8 to 1.5 mm2
Conductor colors: Blue, Brown, Green/Yellow
9 to 12 mm
6 to 9 mm
22
Summary of Contents for SOLYS Gear Drive
Page 1: ...SOLYS2 Sun Tracker SOLYS Gear Drive Sun Tracker Instruction Manual...
Page 2: ...2...
Page 4: ...4...
Page 6: ...6...
Page 10: ...10...
Page 39: ...Figure 3 13 Adjustment of the sun sensor 21 0 5 mm 3x 39...
Page 61: ...6 Insert the Ethernet cable Contacts on this side 61...
Page 78: ...78...
Page 80: ...80...
Page 86: ...86...
Page 88: ...88...
Page 112: ......