D.6 Reset commands
Reset commands restart the suntracking process (track), the instrument (warm) or clear suntracking parameters and restart the
instrument (cold).
D.6.1 Suntracking reset
Description
The command ‘.reset track;’ will restart the suntracking process while the instrument’s internal computer keeps running.
Argument (none)
Example
Request
.reset track;
Response
.ok;
D.6.2 Warm reset
Description
The command ‘.reset warm;’ will restart the instrument.
Argument (none)
Example
Request
.reset warm;
Response
.ok;
D.6.3 Cold reset
Description
The command ‘.reset cold;’ will revert all suntracking configuration parameters to defaults and restart the instrument. This
command has the same effect as the ‘.clear track’ command, see paragraph D.5.1.
Argument (none)
Example
Request
.reset cold;
Response
.ok;