:MARKer:MODE
Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual | WAVeform | BINary | HEX}
The :MARKer:MODE command sets the cursors mode:
• OFF — removes the cursor information from the display.
• MANual — enables manual placement of the X and Y cursors.
If the front-panel cursors are off, or are set to the front-panel Hex or Binary
mode, setting :MARKer:MODE MANual will put the cursors in the front-panel
Normal mode.
• MEASurement — cursors track the most recent measurement.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement source
(:MEASure:SOURce). Setting the measurement source remotely always sets the
marker sources.
• WAVeform — the Y1 cursor tracks the voltage value at the X1 cursor of the
waveform specified by the X1Y1source, and the Y2 cursor does the same for the
X2 cursor and its X2Y2source.
• BINary — logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed in the Cursor sidebar dialog in binary.
• HEX — logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed in the Cursor sidebar dialog in hexadecimal.
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN | WAV | BIN | HEX}
See Also
•
"Introduction to :MARKer Commands"
•
•
•
•
•
•
•
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...