362
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
17
:FFT Commands
:FFT:WINDow
Command Syntax
:FFT:WINDow <window>
<window> ::= {RECTangular | HANNing | FLATtop | BHARris}
The :FFT:WINDow command allows the selection of four different windowing
transforms or operations for the FFT (Fast Fourier Transform) function.
The FFT operation assumes that the time record repeats. Unless an integral
number of sampled waveform cycles exist in the record, a discontinuity is created
between the end of one record and the beginning of the next. This discontinuity
introduces additional frequency components about the peaks into the spectrum.
This is referred to as leakage. To minimize leakage, windows that approach zero
smoothly at the start and end of the record are employed as filters to the FFTs.
Each window is useful for certain classes of input signals.
• RECTangular — useful for transient signals, and signals where there are an
integral number of cycles in the time record.
• HANNing — useful for frequency resolution and general purpose use. It is good
for resolving two frequencies that are close together, or for making frequency
measurements. This is the default window.
• FLATtop — best for making accurate amplitude measurements of frequency
peaks.
• BHARris (Blackman-Harris) — reduces time resolution compared to the
rectangular window, but it improves the capacity to detect smaller impulses
due to lower secondary lobes (provides minimal spectral leakage).
Query Syntax
:FFT:WINDow?
The :FFT:WINDow? query returns the value of the window selected for the FFT
function.
Return Format
<window><NL>
<window> ::= {RECT | HANN | FLAT | BHAR}
See Also
•
•
•
•
•
•
•
•
•
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...