1008
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
33
:TRIGger Commands
The following trigger types are available (see
• Edge triggering— identifies a trigger by looking for a specified slope and
voltage level on a waveform.
• Nth Edge Burst triggering— lets you trigger on the Nth edge of a burst that
occurs after an idle time.
• Pulse width triggering— (:TRIGger:GLITch commands) sets the oscilloscope to
trigger on a positive pulse or on a negative pulse of a specified width.
• Pattern triggering— identifies a trigger condition by looking for a specified
pattern. This pattern is a logical AND combination of the channels. You can also
trigger on a specified time duration of a pattern.
• TV triggering— is used to capture the complicated waveforms of television
equipment. The trigger circuitry detects the vertical and horizontal interval of
the waveform and produces triggers based on the TV trigger settings you
selected. TV triggering requires greater than ¼ division of sync amplitude with
any analog channel as the trigger source.
• USB (Universal Serial Bus) triggering— will trigger on a Start of Packet (SOP),
End of Packet (EOP), Reset Complete, Enter Suspend, or Exit Suspend signal on
the differential USB data lines. USB Low Speed and Full Speed are supported
by this trigger.
Reporting the Setup
Use :TRIGger? to query setup information for the TRIGger subsystem.
Return Format
The return format for the TRIGger? query varies depending on the current mode.
The following is a sample response from the :TRIGger? query. In this case, the
query was issued following a *RST command.
:TRIG:MODE EDGE;SWE AUTO;NREJ 0;HFR 0;HOLD +60.0000000000000E-09;
:TRIG:EDGE:SOUR CHAN1;LEV +0.00;SLOP POS;REJ OFF;COUP DC;
:TRIG:ZONE:STAT 0
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...