300
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
12
:DEMO Commands
:DEMO:FUNCtion
Command Syntax
:DEMO:FUNCtion <signal>
<signal> ::= {SINusoid | NOISy | PHASe | RINGing | SINGle | AM | CLK
| GLITch | BURSt | MSO | RUNT | TRANsition | RFBurst
| SHOLd | LFSine | FMBurst | ETE | CAN | LIN | UART
| I2C | SPI | I2S | CANLin | ARINc | FLEXray | MIL
| MIL2 | NMONotonic | DCMotor | HARMonics | COUPling
| CFD | SENT | KEYSight}
The :DEMO:FUNCtion command selects the type of demo signal:
Demo Signal
Function
Demo 1 Terminal
Demo 2 Terminal
SINusoid
5 MHz sine wave @ ~ 6 Vpp, 0 V
offset
Off
NOISy
1 kHz sine wave @ ~ 2.4 Vpp, 0.0 V
offset, with ~ 0.5 Vpp of random noise
added
Off
PHASe
1 kHz sine wave @ 2.4 Vpp, 0.0 V
offset
1 kHz sine wave @ 2.4 Vpp, 0.0 V
offset , phase shifted by the amount
entered using the
RINGing
500 kHz digital pulse @ ~ 3 Vpp,
1.5 V offset, and ~500 ns pulse width
with ringing
Off
SINGle
~500 ns wide digital pulse with
ringing @ ~ 3 Vpp, 1.5 V offset
Press the front panel
Set Off
Single-Shot
softkey to cause the
selected single-shot signal to be
output.
Off
AM
26 kHz sine wave, ~ 7 Vpp, 0 V offset
Amplitude modulated signal, ~ 3 Vpp,
0 V offset, with ~13 MHz carrier and
sine envelope
CLK
3.6 MHz clock @ ~2 Vpp, 1 V offset,
with infrequent glitch (1 glitch per
1,000,000 clocks)
Off
GLITch
Burst of 6 digital pulses (plus
infrequent glitch) that occurs once
every 80 µs @ ~3.6 Vpp, ~1.8 V offset
Off
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...