1034
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
33
:TRIGger Commands
:TRIGger[:EDGE]:REJect
Command Syntax
:TRIGger[:EDGE]:REJect <reject>
<reject> ::= {OFF | LFReject | HFReject}
The :TRIGger[:EDGE]:REJect command turns the low-frequency or high-frequency
reject filter on or off. You can turn on one of these filters at a time.
• The high frequency reject filter adds a 50 kHz low-pass filter in the trigger path
to remove high frequency components from the trigger waveform. Use the high
frequency reject filter to remove high-frequency noise, such as AM or FM
broadcast stations, from the trigger path.
• The low frequency reject filter adds a 50 kHz high-pass filter in series with the
trigger waveform to remove any unwanted low frequency components from a
trigger waveform, such as power line frequencies, that can interfere with proper
triggering.
Query Syntax
:TRIGger[:EDGE]:REJect?
The :TRIGger[:EDGE]:REJect? query returns the current status of the reject filter.
Return Format
<reject><NL>
<reject> ::= {OFF | LFR | HFR}
See Also
•
"Introduction to :TRIGger Commands"
•
•
NOTE
The :TRIGger[:EDGE]:REJect and the :TRIGger[:EDGE]:COUPling selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:COUPling can change the COUPling setting.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...