:MEASure Commands
22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
479
:MEASure:FALLtime
Command Syntax
:MEASure:FALLtime [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:FALLtime command installs a screen measurement and starts a
fall-time measurement. For highest measurement accuracy, set the sweep speed
as fast as possible, while leaving the falling edge of the waveform on the display. If
the optional source parameter is specified, the current source is modified.
Query Syntax
:MEASure:FALLtime? [<source>]
The :MEASure:FALLtime? query measures and outputs the fall time of the
displayed falling (negative-going) edge closest to the trigger reference. The fall
time is determined by measuring the time at the upper threshold of the falling
edge, then measuring the time at the lower threshold of the falling edge, and
calculating the fall time with the following formula:
fall time = time at lower threshold - time at upper threshold
Return Format
<value><NL>
<value> ::= time in seconds between the lower threshold and upper
threshold in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...